SAMPLE DATA
The following data was provided to the Royal Society Chemistry, MicroAnalytical Group as their baseline test data for a performance evaluation exercise.
|
| Sample Run | %C | %H | %N |
| 1 | 65.43 | 6.72 | 8.45 |
| 2 | 65.47 | 6.73 | 8.45 |
| 3 | 65.47 | 6.72 | 8.48 |
| 4 | 65.45 | 6.70 | 8.44 |
| 5 | 65.44 | 6.72 | 8.47 |
| 6 | 65.50 | 6.72 | 8.52 |
| 7 | 65.52 | 6.70 | 8.52 |
| 8 | 65.48 | 6.70 | 8.49 |
| 9 | 65.45 | 6.70 | 8.48 |
| 10 | 65.49 | 6.71 | 8.48 |
| |
| Mean Values | 65.47 | 6.71 | 8.48 |
| Theoretical Values | 65.44 | 6.71 | 8.45 |
| Deviation from theory | 0.03 | 0.00 | 0.03
|
|

C9H11NO2
Benzocaine; Ethyl 4-aminobenzoate; 4-Aminobenzoic acid ethyl ester; Anesthesin
|
This same sample was run recently in another Royal Society of Chemistry, MicroAnalytical Group Perforamnce Evaluation Exercise. The CE440 was placed in positions 1,2 & 3. This is the only independent testing program carried out by any international organisation in the world that we are aware of. Please click here for full unedited RSC report.