CE440 Elemental Analyser
Sample Data

The following data was provided to the Royal Society Chemistry, MicroAnalytical Group as their baseline test data for a performance evaluation exercise.

 

Sample Run %C %H %N
1 65.43 6.72 8.45
2 65.47 6.73 8.45
3 65.47 6.72 8.48
4 65.45 6.70 8.44
5 65.44 6.72 8.47
6 65.50 6.72 8.52
7 65.52 6.70 8.52
8 65.48 6.70 8.49
9 65.45 6.70 8.48
10 65.49 6.71 8.48
Mean Values 65.47 6.71 8.48
Theoretical Values 65.44 6.71 8.45
Deviation from theory 0.03 0.00 0.03

data-diagramC9H11NO2

Benzocaine;
Ethyl 4-aminobenzoate; 4-Aminobenzoic acid ethyl ester; Anesthesin

Performance Evaluation ExcerciseThis same sample was run recently in another Royal Society of Chemistry, MicroAnalytical Group Perforamnce Evaluation Exercise. The CE440 was placed in positions 1,2 & 3. This is the only independent testing program carried out by any international organisation in the world that we are aware of.